Maintenance-5110
Cathode-Ray
Tube
Replacement
The
following
procedure outlines the
removal and
replacement of
the cathode-ray
tube.
Refer to Fig. 4-7
WARNING |
Use care
when
handling a crt. Protective clothing
and
safety glasses
should be worn. Avoid striking it
on any
object which might
cause it to crack or
implode.
When storing a crt, place it in a protective
carton
or set it face
down in
a
protected
location
on a
smooth surface with
a soft mat under the faceplate to
protect it from
scratches.
A.
REMOVAL
1.
Remove the bezel assembly, which
is held in place
with two screws. (The bezel
assembly
includes a snap-in
implosion
shield.)
O.arrn .ismtsly
CRT
2. Remove the crt base cover on the rear panel of the
instrument.
3.
Remove
the crt
base socket.
4.
Loosen
the crt clamp. The crt and neck portion of
the
shield will
be removed
as a unit, and to facilitate
removal, it may be best
to remove all hardware from the
crt
clamp
(bracket
and positioning
screws, and clamp
tightening
hardware).
NOTE
The red and black wires
entering the crt shield are
connected
to the trace-rotation
coil inside the
shield.
They
will
not hamper crt removal
and need
not
be unsoldered.
5. With one hand on the crt faceplate, push on the crt
base
(and neck
shield). Slide the crt and neck
shield
forward, and
pull
the crt out of the instrument from the
front, then
remove the neck shield.
B.
REPLACEMENT
1.
Slide the neck shield onto the crt neck.
2. Make sure the soft plastic crt faceplate supports are
in
place,
then insert
the crt
into the main shield. Before the
crt
is completely inserted, slidethe crt
clamp overthe neck
shield.
3.
With
the crt fully inserted and loose in the shield,
mount
the bezel assembly into place and tighten
the bezel
screws.
1
Fig. 4-7. Replacing
the cathode-ray tube.
21
14.14
4.
Mount
the
crt clamp and
positioning hardware,
temporarily leaving it loose.
5. Position the rear
of the crt (socket end) so there is no
tilt
of
the faceplate in relation to the bezel assembly.
Tighten
the positioning
screws, then tighten the clamp
hardware.
6.
Place the
crt
base socket onto the crt base pins.
Replace the cover.
7. Replacing
the crt will require partial instrument
adjustment. Refer to
the
Adjustment section of this
manual.
4-15
Summary of Contents for 5110
Page 6: ...5110 Fig 1 1 5110 Oscilloscope 2134 01 ...
Page 22: ......
Page 34: ......
Page 42: ......
Page 83: ...5110 I43XII3 2134 15 ...
Page 84: ... i 143 I132 2134 15 BLOCK DIAGRAM ...
Page 85: ... t 2 E I 5110 S LNHNOdWOD BOVJHaiNI ...
Page 88: ...J4 O3 INTERFACE J INTERFACE A NOV 1978 ...
Page 92: ...AUXILIARY BOARD 1143 2134 18 AUXILIARY BOARD ...
Page 95: ...DEFLECTION AMPLIFIERs ...
Page 102: ...LV POWERSUPPLY CALIB COMPONENTS A 4 L V Power Supply circuit board 5110 ...
Page 107: ...SIGNALS OUT OPTION 7 ONLY SI 10 OPT 7 213 26 REV A FE BI979 SIGNALS OUT COPTION T ONLY n ...
Page 114: ......
Page 115: ...5110 OSCILLOSCOPE ...
Page 116: ...FIG 2 MAINFRAME p 5110 OSCILLOSCOPE ...
Page 117: ...1 ...
Page 121: ...o ...
Page 122: ...S 1 5110 OSCILLOSCOPE ...
Page 123: ...13 FIG 4 RACKMOUNT CABINET 5110 OSCILLOSCOPE ...
Page 124: ......