Maintenance
Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual
4- 13
Troubleshooting Procedure
Figures 4--3 through 4--5 are troubleshooting procedure flowcharts. Use them to
troubleshoot an instrument failure. Begin with Figure 4--3.
WARNING.
Before removing the cabinet, disconnect the power cord from the line
voltage source. Failure to do so could cause serious injury or death.
Yes
No
Done.
Power on the
instrument.
Perform the AC Power Supply
troubleshooting procedure
(Figure 4--4).
Yes
No
Perform the Module Isolation
troubleshooting procedure
(Figure 4--5).
Does
the instrument pass
all the
diagnostics?
Does
the instrument
display the power-up
screen?
Figure 4- 3: Primary troubleshooting procedure