Adjustment Procedures
Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual
3- 3
NOTE
. Do not turn any knobs or push any front-panel buttons other than the
Next Step or Previous Step buttons during the adjustment procedure. Doing so
will cause the oscilloscope to abort the adjustment procedure. The oscilloscope
uses the previous calibration constants if the adjustment procedure is aborted or
fails.
The screen does not display the actual oscilloscope settings (such as channel
input impedance, or vertical and horizontal settings) during the adjustment
procedure. The oscilloscope automatically sets the instrument settings, but these
settings may not read out correctly on the display.
During some steps, the instrument may appear to be idle for several minutes
while it is processing information internally.
If the oscilloscope completes all steps in the procedure successfully, a “Pass”
message is displayed and the new calibration constants take effect. If the
oscilloscope does not pass a step, you can repeat the step. To cancel the
calibration procedure, press the
Menu Off
front panel button on the oscilloscope.
This reverts the oscilloscope to the old calibration constants.
Summary of Contents for 4000 Series
Page 8: ...Service Safety Summary vi Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual ...
Page 13: ...Operating Information ...
Page 14: ......
Page 16: ...Operating Information 1 2 Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual ...
Page 17: ...Theory of Operation ...
Page 18: ......
Page 22: ...Theory of Operation 2 4 Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual ...
Page 23: ...Adjustment Procedures ...
Page 24: ......
Page 37: ...Maintenance ...
Page 38: ......
Page 55: ...Mechanical Parts List ...
Page 56: ......