Adjustment Procedures
3- 12
Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual
Problem
Things to try
I’m having trouble with tests that
require 6 V or --6 V
Follow the steps for 6 V and --6 V that appear in Step 6 of
the DC Voltage test.
Ensure that the polarity is correct, by referring to the icon
that appears in the upper--left corner of the calibrator screen.
If the instrument fails a test, go all the way back to the first 6
V test (6 V on Channel 1) and start the sequence over again.
The Sine Signal test failed
If the prompt specified Channel 4, try setting the signal to
Channel 5, instead.