Adjustment Procedures
Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual
3- 11
CAUTION.
Keep the digital probe plugged in to the oscilloscope. Unplugging the
probe may result in aborting the test.
When a Sine Signal prompt appears, complete the following steps:
1.
Push the
Sine
front panel button.
2.
To ensure that the voltage is in direct mode, push the
1 2
5
1.Ø
lower--bezel
button and select
1.0
.
3.
Push the
CHANNEL SELECT
lower--bezel button, and then ensure that
Channel 5 is selected.
4.
Ensure that the red light is illuminated to indicate that the output is on. If
not, push the
ON
button.
5.
On the oscilloscope, push the
OK Do Next Step
right--bezel button.
Refer to the next table for common issues encountered during the calibration
procedure.
Problem
Things to try
The instrument fails a test
Are the channels hooked up properly, routed to the right
channels, and turned on?
Is the signal set to the correct waveform, frequency, and
termination (if applicable to the test)?
Is the output set to
ON
(the red light will be illuminated)?
If you make a correction or find that everything appears to be
correct, click the
Go back a step
button to try the test again
.
I can’t set the amplitude to 0 V,
as specified in the prompt
Ground the signal by pushing the
lower--bezel
button.
I can’t find the parameters that I
need to set
Is the waveform set correctly?
Is the grounding button on? When this is active, you will not
be able to make a channel selection, so in some cases you
might need to temporarily deselect the grounding option
while you make a channel selection.
If you can find the termination settings, click the
CHANNEL
SELECT
lower--bezel button.
I can’t set the voltage to 625 Hz
Set the time marker to 1.6 Ms. This will automatically set the
voltage to 625 Hz.
I can’t set the signal to all
channels
If you are unable to set the signal to all channels, try setting
the signal to Channel 1.
Sine Signal
Troubleshooting
Summary of Contents for 4000 Series
Page 8: ...Service Safety Summary vi Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual ...
Page 13: ...Operating Information ...
Page 14: ......
Page 16: ...Operating Information 1 2 Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual ...
Page 17: ...Theory of Operation ...
Page 18: ......
Page 22: ...Theory of Operation 2 4 Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual ...
Page 23: ...Adjustment Procedures ...
Page 24: ......
Page 37: ...Maintenance ...
Page 38: ......
Page 55: ...Mechanical Parts List ...
Page 56: ......