Adjustment Procedures
Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual
3- 7
12.
Push the
OK Do Factory Calibration
side--bezel button, as shown in
Figure 3--4, to start the adjustment process. Prompts appear on the oscillo-
scope screen to indicate the signal type and the channels to which it should
be connected.
NOTE
. The oscilloscope adjusts itself automatically using the calibrator signal
as a reference. You do not need to make any adjustments.
If you make an error, such as connecting the wrong input signal, you can repeat
the last step by pushing the
Return to Previous Screen
side bezel button.
13.
For each step, refer to the next table to identify the signal type, and then
carefully follow the instructions for that signal type on the specified page.
Signal type
Example prompt
Refer to
Parameters to set
DC Voltage
Apply 0 V DC 50
Ω
ter-
mination signal to 3.
Page 3--8
Waveform type, voltage,
termination, channel
Time Mark
Apply 500 mV Pk--Pk 625
Hz 50
Ω
termination
time mark signal to 3.
Page 3--9
Waveform type, voltage,
frequency, termination,
channel
Deskew
Apply 1.0 V Pk--Pk 1 kHz
50
Ω
termination deskew
signal to 3.
Page 3--10
Waveform type, voltage,
frequency, termination,
channel
Sine Signal
Apply 2.0 V Pk--Pk 164M
50
Ω
termination sine
signal to 3.
Page 3--11
Waveform type, voltage,
frequency, termination,
channel
NOTE
. The oscilloscope screen will not provide you with all of the information
that you need to complete the steps successfully, so it is critical that you follow
the steps outlined in the Completing the Procedure section. Each time a new
prompt appears, refer to the table above to identify the signal type, and then
ensure that you are following every step that is outlined for that specific signal
type.
14.
Continue with the adjustment process until it is complete and you receive the
Pass
notification. You can cancel the process at any time by pushing the
MENU OFF
button.
15.
Complete the performance verification tests to verify that the adjustment
procedure has correctly calibrated the oscilloscope. (See the
Tektronix 4000
Series Technical Reference
.) Incorrect use of calibration equipment can cause
the oscilloscope to pass the Calibration Procedure but fail Performance
Verification.
Summary of Contents for 4000 Series
Page 8: ...Service Safety Summary vi Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual ...
Page 13: ...Operating Information ...
Page 14: ......
Page 16: ...Operating Information 1 2 Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual ...
Page 17: ...Theory of Operation ...
Page 18: ......
Page 22: ...Theory of Operation 2 4 Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual ...
Page 23: ...Adjustment Procedures ...
Page 24: ......
Page 37: ...Maintenance ...
Page 38: ......
Page 55: ...Mechanical Parts List ...
Page 56: ......