Adjustment Procedures
Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual
3- 5
6.
Insert the adjustment tool into the CAL switch hole. Use the tool to push and
hold the CAL switch in while you power on the oscilloscope. Keep holding
the switch until the oscilloscope shows the “Power-on self--test passed”
message with the oscilloscope graticule in the background. This can take up
to a minute.
7.
Push the
Default Setup
button on the front panel of the oscilloscope.
8.
Push the
Utility
button on the front panel of the oscilloscope.
9.
Push the
Utility Page
lower--bezel button. The utility page menu appears, as
shown in Figure 3--2.
Figure 3- 2: Utility page menu
10.
Use the
Multipurpose a
knob to select
Calibration
, as shown in
Figure 3--3.