Calibration
—
Type
3B5
Fig.
6-7. Initial test equipment setup for steps 7—16.
Control
Settings
Type
3B5
Mode
MAN
POSITION
Midrange
DLY'D
SWP MAG
OFF
DELAY
1.00
MANUAL
TIME/DIV
1
ms
VARIABLE
CAL
Trigger
Function
INT-AUTO
LEVEL
Midrange
SLOPE
+
Amplifier
Unit
Position
Midrange
Input
Coupling
DC
Volts/Division
0.5
Volts
Mode
Manual
7.
Adjust
Basic
Timing
a. Test
equipment
setup
is
shown
in
Fig.
6-7.
b.
Connect
the
time-mark
generator
to
the
amplifier
unit
input
connector
through the 50-ohm
BNC
cable
and
a
50-ohm
BNC
termination.
c.
Set the
time-mark
generator
for
one-millisecond
markers.
d.
Position
the
second
marker
to
the
first graticule
line
with
the
POSITION
control.
e. CHECK
—CRT
display
for
one
marker
each division
between
the
first
and
ninth graticule
lines
(see Fig.
6-8A).
NOTE
Unless
otherwise
noted,
use
the
middle
eight
hori
zontal
divisions
when
checking
or
adjusting
timing
(see
Fig.
2-4,
Operating
Instructions
section).
Indicator
Oscilloscope
Focus
Adjust
for
correct
display
Intensity
Visible
Display
Calibrator
10
Volts
Power
On
f.
ADJUST
—CAL
adjustment,
R515
(see
Fig.
6-8B),
for
one
marker
each
division.
g.
INTERACTION—
Check
steps
10
—
21.
(This
adjustment
should
be
checked
and
readjusted
as
necessary
when
the
Type
3B5
is inserted
into
a
plug-in
compartment other
than
the
one
in
which
it
was
calibrated.)
6-10
Summary of Contents for 3B5
Page 4: ...Fig 1 1 Type 3B5 Automatic Programmable Time Base unit Type 3B5...
Page 15: ...Operating Instructions Type 3B5 TYPE 3B5 CONTROL SET UP CHART Fig 2 2 Control set up chart 2 7...
Page 48: ...CO I o Fig 3 13 Delay and Timing Circuit logic block diagram Circuit Description Type 3B5...
Page 61: ...GO i GO GO Fig 3 22 Seek Ciicuit Logic block diagram Circuit Description Type 3B5...
Page 70: ...u k KJ Fig 3 29 Circuit conditions for Manual Mode operation Circuit Description Type 3B5...
Page 71: ...w K w Fig 3 30 Circuit conditions for Seek Mode operation Circuit Description Type 3B5...
Page 72: ...w I u U Fig 3 31 Circuit condition for External Mode operation Circuit Description Type 3B5...
Page 88: ...Maintenance Type 3B5 Fig 4 9 Location of components on Logic Card 4 14...
Page 89: ...u Oi Fig 4 10 Location of components on Counter Card Maintenance Type 3B5...
Page 92: ...NOTES I...
Page 104: ...NOTES...
Page 106: ...Calibration Type 3B5 Fig 6 1 Recommended calibration equipment...
Page 160: ......
Page 176: ...J400 RtADOUT BOARD 3B5 PLUG IN A READOUT...
Page 182: ...397 R E A D O U T B O A R D 10 6b READOUT BOARD...
Page 184: ...FIG 1 FRONT SWITCHES TYPE 3B5 AUTOMATIC PROGRAMMABLE TIME BASE...
Page 185: ...FIG 2 CHASSIS REAR 3 GS to TYPE 3B5 AUTOMATIC PROGRAMMABLE TIME BASE...
Page 186: ...OPTIONAL ACCESSORIES...