Characteristics
—
Type
3B5
HORIZONTAL
DEFLECTION
SYSTEM
Characteristic
Performance Requirement
Supplemental
Information
Sweep
Rates
Manual
Mode
5
seconds/division to
10 nanoseconds/division
in
27
calibrated
steps.
Sweep
rate
indicated
on
readout
panel
Selected
by
front-panel
MANUAL
TIME/
DIV
switch. Steps
in
1-2-5
sequence
Seek
Mode
5
seconds/division
to
3.1
microseconds/divi-
sion
in
24
calibrated
sleps.
Sweep
rate
indi
cated
on
readout panel
Automatically
selected
when
seek
com
mand
is
received.
Steps
in
1-2-5
sequence
External
Mode
5
seconds/division
to
10
nanoseconds/division
in
27
calibrated
steps.
Sweep
rate
indicated
on
readout
panel
Selected
by
external
program
connected
to
PROGRAM
connector.
Steps
in
1-2-5
sequence
Sweep
Accuracy
(at
25°
C,
±5°)
1s
to
10
ns
Within
±3% of indicated sweep
rate
VARIABLE control
fully
clockwise
(UN
CAL
readout
off)
and
DLY'D
SWP
MAG
switch set
to
OFF (MAG'D
SWP
readout
off).
10, 20 and
50
ns
accuracy must
be
checked
in
accordance
with
Performance
Check
5
s
and
2
s
Within
±5% of
indicated
sweep
rate
Variable Sweep
Rate
Uncalibrated
sweep
rate
to
2.5
times,
or
greater,
the
indicated
sweep
rate.
s/DIV
read
out
turns
off
and
UNCAL
readout
turns
on
when
VARIABLE
control is
not
set
to
CAL
posi
tion
Slowest
sweep
rate
12.5
seconds/divi
sion, or greater,
in
5
s
position
Sweep
Length
DLY
’
D
SWP
MAG
at
OFF
10.3
to
11.3
divisions
Measured
at
1
ms/DIV
sweep
rate
DLY'D
SWP
MAG
on
12
to
15
divisions
Delayed
Sweep
Magnification
Basic
sweep
rates
between
5
s
and
1
μs
can
be
magnified
with
the
Dela
yed
Sweep
Magnifier.
Calibrated
DELAY
control
selects
portion to be
magnified.
Basic
sweep
rates
between
5
s
and
10
μs
can
be
magnified
100
times;
basic
sweep
rates
between
5
s
and
1
μ
s
can
be
magnified
10
times
Readout
panel
indicates
correct
magni
fied
sweep
rate
Magnified
Sweep
Accuracy
(at
25°
C,
±5°)
1
S
to
1
JUS
Within
±
3%
of sweep
rate indicated
on
read
out
panel
VARIABLE control
fully
clockwise
(UN-
CAL
readout off)
5
s and
2
s
Within
±5%
of
sweep
rate
indicated
on
read
out
panel
Calibrated
Delay
Range
Continuous
from
0 to
10
divisions from
the
start
of
the
trace
Delay
Accuracy
Within
±1.5%
of fu
l
l scale
at
1
ms/DIV.
Within ±3%
of
full
scale
at
all
other
sweep
rates
Calibrated
at
1
ms/DIV.
Exclude
first
0.2
division
of
display.
Includes
incre
mental
DELAY
dial
linearity
Incremental
DELAY
Dial Linearity
Within
±0.4% of
full
scale
Exclude
first
0.2
division
of
display
Delayed
Sweep
Magnifier
Jitter
3
parts
or less
per
10,000
Equal
to
0.3
division
or less
when meas
ured with
x1
00
magnification
at
1
ms/
DIV
basic
sweep
rate
Output
DC
Level
4-180
volts,
±10
volts
Beam
positioned
to CRT
electrical
center
SEEK MODE
Cycles
Per
Sweep
Between
two and
six
cycles
with
nominal
CYCLES/SWEEP
setting obtained
in
Calibra
tion
procedure
Automatically
selected
when
seek
com
mand is received.
Front-panel
CYCLES/
SWEEP adjustment
provides
range
of
about
0.5 to
3
times
this
nominal setting
Automatic
Seek
Cycling
Time
One
to
two seconds
When
seek
command
is
continuous
Automatic
Seeking
Time
500
milliseconds
or
less with 10
MHz
sine
wave
input
and
Trigger
Function
switch
set
to
INT-AUTO
Frequency
Range
for
Seek
Mode
Operation
30
Hz
to
20
MHz
Seek
Mono
Pulse
Width
1.0
to
1.8
seconds
1-2
Summary of Contents for 3B5
Page 4: ...Fig 1 1 Type 3B5 Automatic Programmable Time Base unit Type 3B5...
Page 15: ...Operating Instructions Type 3B5 TYPE 3B5 CONTROL SET UP CHART Fig 2 2 Control set up chart 2 7...
Page 48: ...CO I o Fig 3 13 Delay and Timing Circuit logic block diagram Circuit Description Type 3B5...
Page 61: ...GO i GO GO Fig 3 22 Seek Ciicuit Logic block diagram Circuit Description Type 3B5...
Page 70: ...u k KJ Fig 3 29 Circuit conditions for Manual Mode operation Circuit Description Type 3B5...
Page 71: ...w K w Fig 3 30 Circuit conditions for Seek Mode operation Circuit Description Type 3B5...
Page 72: ...w I u U Fig 3 31 Circuit condition for External Mode operation Circuit Description Type 3B5...
Page 88: ...Maintenance Type 3B5 Fig 4 9 Location of components on Logic Card 4 14...
Page 89: ...u Oi Fig 4 10 Location of components on Counter Card Maintenance Type 3B5...
Page 92: ...NOTES I...
Page 104: ...NOTES...
Page 106: ...Calibration Type 3B5 Fig 6 1 Recommended calibration equipment...
Page 160: ......
Page 176: ...J400 RtADOUT BOARD 3B5 PLUG IN A READOUT...
Page 182: ...397 R E A D O U T B O A R D 10 6b READOUT BOARD...
Page 184: ...FIG 1 FRONT SWITCHES TYPE 3B5 AUTOMATIC PROGRAMMABLE TIME BASE...
Page 185: ...FIG 2 CHASSIS REAR 3 GS to TYPE 3B5 AUTOMATIC PROGRAMMABLE TIME BASE...
Page 186: ...OPTIONAL ACCESSORIES...