34
EMI (Electromagnetic Interference)
WARNING: Keep the patient under strict supervision while the shock is being
delivered. It is possible, although improbable, that electromagnetic signals
radiating from sources outside of the AED can lead to inaccurate values. Do not
rely exclusively on the measurements provided by the AED for your assessment
of the patient.
WARNING: It is possible that any device designed for radio frequency transmission
and other nearby sources of electrical disturbance can cause
interruptions in AED operation.
WARNING: It is possible, although improbable, that the use of power on/off relay
switches can cause the AED to malfunction. Do not use the AED with
electrocauterising or diathermic devices, or in similar environments.
This AED was tested and has been proven to be compliant with the standards required for
medical devices by IEC60601-1-2, and by Directive 93/42/EEC regarding medical devices.
These limits are established to provide suitable protection against any harmful interference
in traditional medical facilities.
Nonetheless, due to the proliferation of devices that transmit radio frequencies and other
sources of electrical interference in hospital facilities (electro-surgical devices, mobile
phones, two-way mobile radios, electrical devices and HD television sets), high levels of this
interference due to the excessive vicinity or intensity of a source, can cause an interruption
in AED operation.
WARNING: The AED is designed for use in environments where the signal can be
affected by electromagnetic interference. During such interference, the
measurements may be inexact or it may seem that the AED is not working
correctly.
Defective AED operation can be illustrated by irregular measurements, interruptions in
operation or other malfunctions. If this should happen, inspect the area to determine the
source of the disturbance. Perform the following actions to attempt to eliminate any
malfunctions.
Turn the nearby devices off and back on again to identify the interfering device.
Redirect or reposition the interfering device.
Increase the distance between the interfering device and the affected device.
The AED generates, uses and can release radio-frequency energy. If the AED is not
installed and used in compliance with these instructions, the AED can cause harmful
interference with other nearby devices.
If you require assistance contact qualified staff or your local supplier.
Obtaining Technical Assistance
To receive information and assistance, or to order the AED maintenance manual, contact
your local supplier. The maintenance manual provides the necessary information to
qualified staff or your local dealer on AED maintenance.
For technical assistance, please contact the store that invoiced the device, or contact us
directly at:
TECNO-GAZ S.P.A.
Tel. +39 0521 8380 - Fax +39 0521 833391
MEDILINE ITALIA S.r.l.
Tel. +39 0522 942997 – Fax +39 0522 944800
e-mail : [email protected]