17
Battery Status Reading
The new battery life is provided as set forth below;
- Storage life (in the original package): 2 years from the production date, if stored and
maintained according to the methods described in the operator manual.
- Battery life in standby (installed in the AED): 5 years from the production date, if stored
and maintained
according to
the methods described in the operator manual.
- Discharged: A minimum of 200 electrical shocks (not counting the CPR period between
the defibrillation treatment) or 10 hours of operation at an ambient temperature of 20 °C.
TECNO-GAZ suggests replacing the used battery, even if it has only been used once.
Notes: After 200 electrical shocks, a voice message will play ““Low battery, insert fresh
battery” (Battery discharged, change the battery)” .
Notes: Due to the physical measurements of the battery compartment, it is necessary to use
batteries supplied by TECNO-GAZ only. Using other types of spare batteries may
damage the AED and cause the limited warranty to lapse.
When you are working on the batteries, the battery status reading reports the battery level.
See Table 4.
Table 4. Battery Status Icon
Battery Status Icons
Battery Status
Fully charged battery
(it is possible to perform
≤
200 electrical shocks
or 10 hours of operation/defibrillation)
Battery In standard use
Used battery (
≤
9 electrical shocks)
it is necessary to order a new battery, code. 1Z12A0002
Discharged, missing or disconnected battery
(no electrical discharge possible)
If you hear the voice message saying "Batteria scarica, sostituire la batteria (Battery
discharged, change the battery)" when the AED is on or in use, the AED has 9 electrical
shocks left. If the last bar on the level reading is not visible, the buzzer will sound twice and
the device will switch off automatically. Change the battery immediately.