19
Self-test when the Batteries are Inserted
When the battery is installed or changed, the AED automatically runs the battery insertion
self-test. Once the battery insertion self-test is complete, the AED will play the “Unità ok
(unit ok)” voice message, the status indicator will display “O” and the AED will automatically
switch off. If the battery insertion self-test is not completed correctly, the AED will play the
following voice message: “Unità non funzionante (unit not operating)” and the status
indicator will display “X”. If the AED is not working correctly, do not use it. Rather, seek the
assistance of qualified personnel or your local supplier.
It is also possible to skip the battery insertion self-test through the following procedure
Closed: skip by opening the cover.
Open: skip by pressing the shock button.
When you have completed this procedure the AED will run the power-on self-test, exactly
as it does when the user turns the AED on.
Notes: The self-test is not capable of determining whether the battery and the electrodes
connected to the AED have expired. Remember to check the expiration date of the
electrodes and battery life in standby, on a regular basis.