16
a)
b)
1
c)
2
3
To disconnect the battery
To connect the battery
1. AED
connector
2. Hook
3. Handle/battery
connector
Figure 5. Changing the handle/battery -Top Panel
To disconnect the battery:
a) Press the handle/battery removal button, pull upwards, then disconnect the
handle/battery.
To connect the battery:
b) Tilt the handle/battery and hold the handle/battery removal button and the AED
connector perpendicular, as illustrated in Figure 5.
Connect the AED and the handle/battery using the hook, as illustrated in the first
figure of Figure 5.
c) When the handle/battery is connected to the AED by the hook, connect the
handle/battery connector and the hook on the AED, as illustrated in the second figure
of Figure 5.
d) When both parts of the connectors are hooked up correctly, the device will make a
tinkling sound.
The AED uses disposable batteries. Used batteries must be replaced with new batteries.
Before turning on the AED with fully discharged batteries, change the batteries. When the
new battery is installed, the AED will switch on automatically and start the battery insertion
self-test. The AED may switch off when the battery insertion self-test is finished.