18
Self-test
Before using the AED, make sure that the AED is working correctly and that
it is safe to use, as described below.
WARNING: If the self-test is not carried out correctly, do not attempt to use the
AED.
ATTENTION: When the power supply is connected, the AED automatically runs the
self-test, which checks the AED circuits and functions. As the Power On Self Test
(POST) is running, make sure that the AED status indicator is on. If the AED status
indicator is not working correctly, do not use the AED. Rather, seek the assistance of
qualified personnel or your local supplier.
Running the Power On Self-Test (POST)
1. Turn the AED on from the
Scrolling switch
(open the door)
2. The AED will automatically run the Power On Self Test (POST).
3. If the AED detects any problems during POST, the status indicator will display “X”.
Contact qualified assistance staff or your local supplier for assistance.
4. Once you have run the POST correctly, the AED will play this voice message: “Unità ok
(unit ok)” and the status indicator will display “O”.
5. Turn the AED off by closing the cover.
Automatic Self-test
The AED is programmed to run an automatic self-test on a daily basis.
The self-test will run automatically and does not require user control. When an error is
detected, the status indicator displays “X”.
The self-test tests the AED and makes sure that the basic functions are running properly.
Daily self-test: MCU and Memory Integrity (RAM, ROM), Battery capacity, ECG
algorithm analysis.
Weekly self-test: Low energy waveform circuit test (2J), ECG circuit test, in addition
to the daily self-test.
Monthly self-test : High energy waveform circuit test (50J) in addition to the monthly
self-test.
Notes: When the battery is low, the status indicator displays “X”. The battery status indicator
also displays “X” when the battery is changed. Seek the assistance of qualified
personnel or your local supplier.
Notes: The self-test is not capable of determining whether the battery and the electrodes
connected to the AED have expired. Remember to check the expiration date of the
electrodes and battery life in standby, on a regular basis.