14
Installation step 2
Make sure the AED is working correctly.
1. Change the patient mode by moving the Scrolling button to the right or to the left,
depending on whether it is an adult or paediatric patient.
2. Turn the AED on by moving the
Scrolling button
to the right and opening the cover,
making sure that it is possible to hear the voice messages.
“Unità ok (unit ok)”
“Elettrodi adulti (adult electrodes)” or “Elettrodi pediatrici (paediatric electrodes)”
3. Make sure that the status indicator reading of “O” is clearly visible.
4. Turn the AED off by closing the cover.
5. Close the Cover making sure to position the defibrillation electrodes inside the AED.
Notes: When you change the AED battery, the self-test will run automatically. When the
self-test is over, make sure you hear the voice message saying "Unità ok (unit ok)"
and make sure the AED is off.
Installation step 3
Put the AED back in its Soft Case.
Installation step 4
Place the AED in an attended, safe and visible location.
Notes: The location/storage specifications are not the same in different countries. Seek
the assistance of qualified personnel or your local supplier.
The AED must be kept in an easily accessible location. Place the device near a telephone
so that the rescuer can call the emergency medical services and get the AED without
wasting any time. Some important points for storage:
Store the AED in a position that is easy to access.
Do not lock the selected storage area for the AED with a key.
Store the AED in a clean, dry place.
Install the AED according to the ambient operating conditions described herein.
Set things up so that the AED is accessible at all times.
Train and Inform any possible users of the location where the AED is kept.