62
Message
Explanation
LOAD
ADC plus over
−LOAD
ADC minus over
FULL
Display plus over (greater than maximum display value)
−FULL
Display minus over (less than minimum display value)
OVER FULL
Input is exceeding maximum input range (5.0 mV/V)
−OVER FULL
Input is less than minimum input range (−5.0 mV/V)
ZERO OVER
Zero balancing range exceeds regulated values
ZERO ERROR
Zero-adjusting failed within specified time
OUTPUT CAL OVER
Sensor output exceeds calibration range
OUTPUT CAL SHORT
Sensor output does not achieve calibration range
MINUS INPUT
Sensor input is negative
TEDS READ ERROR
A valid TEDS sensor is not connected
PARAMETER ERROR
Irregular setting value exists
R.O. SET OVER
Rated output exceeds the set range (greater than 5.0 mV/V)
R.O. SET SHORT
Rated output below the set range (less than 0.3 mV/V)
ZERO LIMIT OVER
Digital Zero Limit exceeded
ERROR
An error has occurred
DA OVER
DA output is outside output range
DA -OVER
DA output is outside output range
SYSTEM ERROR
A system error has occurred
INVALID OPERATION
Operation is invalid
MEMORY FULL
Memory for saving data is full. Delete data or enable overwriting using "System
settings 2".
Zero Balancing
Please wait a while
Executing Digital Zero
Please wait a while
Reading TEDS data
Please wait a while
WRITING DATA
Do not operate the unit until writing completes.
13. Error message list