28
4-10-2. Condition Setting
Item
Settings
Format
Unit
display
Default
value
Setting range/options/
operation
Cal.
Value
Lock
Setting
value
lock
Setting
value
memory
Filter
Select Num. of Moving
Avg.
Options
OFF
OFF, 16, 32, 64, 128, 256, 512,
1024, 2048
Motion Detect
Time
Input
Seconds
1.5
0.0 to 1.5
Width
Input
Set unit
000.05
00000 to 00999
Zero Tracking
Time
Input
Seconds
0.0
0.0 to 9.9
Width
Input
Set unit
000.00
00000 to 00999
Digital Zero Offset
Input
Set unit
000.00
−19999 to 19999
Digital Zero
Digital Zero Limit Value Input
Set unit
999.99
00000 to 99999
Clear Digital Zero
Options
Skip
Execute, Skip
Select Data Output
Options
Displayed
output
Hold value is output in coor-
dination with display,
input is output as is
4-10-3. Comparison Setting
Item
Setting
Format
Unit
display
Default
value
Setting range/options
Cal.
Value
Lock
Setting
value
lock
Setting
value
memory
Comp. Value Setting
High limit input (HI)
Input
100.00
−99999 to 99999
Low limit input (LO)
Input
10.00
−99999 to 99999
Comp. Mode Setting
Options
Compare
Always
Compare Always,
Compare During Stable,
Compare Except Nearly Zero,
Compare During Stable
Except Nearly Zero,
Compare During Held,
Disable comparison
Hysteresis
Input
Set unit
000.00
00000 to 99999
Nearly Zero
Input
Set unit
001.00
00000 to 09999
4-10-4. Graph settings
Item
Setting
Format
Unit
display
Default
value
Setting range/options
Cal.
Value
Lock
Setting
value
lock
Setting
value
memory
Time axis setting (X)
Options
s
0.25
0.25, 0.5, 1.0, 5.0, 10.0, 30.0
Indicator value axis
setting (Y)
Options
Auto
Auto, Manual
Displayed high limit
value
Input
110.00
−99999 to 99999
Displayed low limit
value
Input
000.00
−99999 to 99999
Trigger mode
Options
OFF
OFF, ON
Trigger
Trigger polarity
Options
SLOPE
2
SLOPE
2
, SLOPE
x
Trigger level
Input
50.00
−99999 to 99999
Trigger position
Options
%
50
0, 10, 20, 30, 40, 50, 60, 70,
80, 90, 100
4. Settings