![Teac TD-01 Portable Owner'S Manual Download Page 40](http://html1.mh-extra.com/html/teac/td-01-portable/td-01-portable_owners-manual_827106040.webp)
40
6-3. Zero Tracking
This function automatically tracks drift and other gradual changes to the zero point.
With Zero Tracking, when the zero point moves no more than the set width, the indicator value is automatically adjusted to zero each
set time interval.
This only functions when Cal. Value Lock is ON and Enable Digital Zero is ON.
Zero Tracking will not function if its Time is set to 0.0 seconds and its Width is set to 0.0.
0
Indicator value
Time
Zero Tracking Width
Zero Tracking Time
Zero Tracking active
Zero Tracking active
6-3-1. Time
6-3-2. Width
6. Condition Setting