ADJUSTMENT OF AMPLITUDE
Low frequencies with high amplitudes can cause discomfort, skin irritations or even burns,
especially if inadequate electrodes or worn self adhesive pads are used.
F-SCAN2 can generate square wave signals with amplitudes of up to 27Vpp with the
potentiometer fully open.
WE STRONGLY RECOMMEND THAT THE AMPLITUDE BE SET CAREFULLY PRIOR
TO A THERAPY WITH LOW FREQUENCIES.
This can be done as follows:
Turn the knob of the potentiometer counter clockwise to ‘0’.
Attach an application cable to the ‘OUTPUT” connector and to the electrodes to be used.
Connect the power supply to an outlet, attach the plug to the socket POWER and set the
switch in the back panel to ON.
The ‘Start-up’ window is displayed. Touch the tab THERA.
REMOVE THE ELECTRODES UNTIL READY FOR TREATMENT. MAKE SURE NOT TO
CHANGE THE SETTING OF THE POTENTIOMETER FOR THIS USER.
Touch “1 2 3“ to display the numerical block.
Input 100(Hz) and confirm with ‘E’.
Select the square wave signal.
HAND THE ELECTRODES TO THE USER.
TURN THE DIAL OF THE POTENTIOMETER SLOWLY
CLOCKWISE (UP) UNTIL THE USER REPORTS
‘FEELING THE CURRENT FLOW’ - OR ‘A SLIGHT
VIBRATION IN HIS HANDS’ - WHICH HE CAN EASILY
TOLERATE.