TP950 Instrument Manual
Rev. J
10
●
Chapter 2: Installation Requirements and Maintenance
1. Installation Conditions
To ensure optimum performance and long and stable product life, observe the following
precautions when selecting an installation location.
Note that installing the system in a location that does not meet the requirements listed below
can result in malfunctions and/or system breakdown. Replacement of components and
system repairs for any deterioration in performance or damage caused by failure to adhere
to the following conditions will not be covered by warranty.
1. Install system in a comparatively dust-free location.
2. Avoid locations subject to corrosive or flammable gases.
3. Avoid locations subject to either strong shocks or continuing low levels of vibration.
4. Install the system away from direct sunlight.
5. Install the system as far away as possible from electrical equipment that generates
strong magnetic fields, electrical fields and high frequencies. Avoid locations near
equipment that generate noise caused by electrical motors and transformers.
6. Choose a flat location capable of withstanding the design weight of the system.
7. The installation environment must be within 15 to 30ºC temperature range and a 20
to 80% relative humidity range.
8. Avoid locations subject to sharp temperature fluctuations, exhaust from
air-conditioning systems and condensation caused by sharp humidity fluctuations.
9. Avoid blocking air vents located on the right and left sides, front and rear of the
system.
10. Choose an operating location with good ventilation.
11. Do not operate the system in a cold room.
12. Select a room that can provide sufficient power (frequency, voltage and current) for
system operation.
13. Do not attempt to move the system while holding it by the lid handle or the control
panel, as this can damage to key components. Doing so can also result in personal
injury due to part breakage resulting in the system falling down.