Table IX. Problem: Apparent low sample concentration.
Possible cause
Solution
Syringe thumbscrew is
loose
Tighten the syringe thumbscrew until finger tight. The thumbscrew is located at
the bottom of the syringe in the Syringe Pump Unit.
Tip is plugged
See Table VI above.
Air bubble in the syringe
path
Prime the syringe path by performing the Daily Warmup.
Microsolenoid valve is
leaky
See “Leaking inlet valve” (Table VI) above.
Syringe valve is blocked
or leaky
See “Leaking inlet valve” (Table VI) above.
Low liquid level in the
Water Container
Check the level of the system liquid in the Water Container. Make sure that the
end of the tubing is submerged in the water. Add deionized, degassed water as
needed (See “Powering on the System” in Section VI.A).
Table X. Problem: System stalls because the syringe does not move.
Possible cause
Solution
No power
Inspect the power cables and connections.
Syringe is not initialized
Cycle the power on the Syringe Pump Unit and restart the computer.
Obstruction
Verify that the syringes are not obstructed.
A.
Technical Support
If you require technical support, please contact your authorized TBUSA service technician, or contact us
directly at:
.
Co
ntact Us
Customer Service/Ordering
Technical Support
tel: 800.662.2566 (toll-free)
tel: 800.662.2566 (toll-free)
fax: 800.424.1350 (toll-free)
fax: 800.424.1350 (toll-free)
web:
takarabio.com
web:
takarabio.com
e-mail:
e-mail:
Notice to Purchaser
Our products are to be used for research purposes only. They may not be used for any other purpose, including, but not limited to, use in drugs,
in vitro
diagnostic
purposes, therapeutics, or in humans. Our products may not be transferred to third parties, resold, modified for resale, or used to manufacture commercial products or to
provide a service to third parties without prior written approval of Takara Bio USA, Inc.
Your use of this product is also subject to compliance with any applicable licensing requirements described on the product’s web page at
takarabio.com
. It is your
responsibility to review, understand and adhere to any restrictions imposed by such statements.
©2018 Takara Bio Inc. All Rights Reserved.
All trademarks are the property of Takara Bio Inc. or its affiliate(s) in the U.S. and/or other countries or their respective owners. Certain trademarks may not be
registered in all jurisdictions. Additional product, intellectual property, and restricted use information is available at
takarabio.com
.
This document has been reviewed and approved by the Quality Department.