SMARTer™ ICELL8® cx Single-Cell System User Manual
(061918)
takarabio.com
Takara Bio USA, Inc.
Page 23 of 35
2.
Click the [Dispense cells] button to initiate dispense.
3.
The vacuum holding the chip will be on automatically, and the system will check the vacuum confirm
that the chip sits flat. If the chip is seated properly, the vacuum radio button will turn green. If the
chip is not secured, the button will be red.
Figure 18. Vacuum status confirmation.
4.
If the chip is not secured, you can take out the chip and restart the process by clicking the “Dispense
cells” button again.
5.
The system will ask for confirmation on source plate and chip loading. Confirm by clicking “OK”.
The dispense step lasts roughly 13 minutes.
Figure 19. Confirm plate and chip loading.
6.
Manually remove the sealing film. Click [Done] in the
Remove seal
window to confirm.
Figure 20. Confirm seal removal.