SMARTer™ ICELL8® cx Single-Cell System User Manual
(061918)
takarabio.com
Takara Bio USA, Inc.
Page 13 of 35
Environmental Controller
The ICELL8 cx unit includes a Proportional-Integral (PI) controller board that monitors the enclosure
temperature, relative humidity (RH) and chip temperature. It adjusts the RH and chip temperature at or near dew
point temperature to minimize evaporation during sample dispensing.
Wash Container
The Wash Container contains 0.2% sodium hypochlorite solution which is used during the tip cleaning steps of
the dispensing protocol to prevent cross-contamination. The sodium hypochlorite solution is pumped from the
Wash Container to the wash station through the pump box.
A.
ICELL8 cx Specifications and Lab Requirements
Table II. ICELL8 cx specifications and lab requirements.
Category
Specification
Dispense volume
50 nl or 100 nl per nanowell
Software
SMARTer ICELL8 cx Software
Laptop computer
Windows 7, 2 GB memory, 120 GB storage, 1 GB network adapter, USB ports
for memory sticks
Power requirements
(for different
electrical grid types)
120 VAC/60 Hz mains:
four 120 V, 15 or 20A circuits
220
–240 VAC/50 Hz mains:
four 10A outlets
100 VAC/50
–60 Hz mains:
four 15A outlets
Fuses
Dispenser; 4AH, 250 V
Environmental
conditions
Ambient temperature: 15
–30°C
Relative humidity, non-condensing: 30
–70%
Altitude: <2,000 m from sea level
Pollution degree: 2 or less
Dimensions
Desktop Computer Tower: 7
” W x 14.5” H x 17.5” D (18 cm x 37 cm x 44 cm)
ICELL8 cx Module: 42
” W x 22” H x 22.5” D (107 cm x 56 cm x 57 cm)
Bench space
Bench space including clearance for ICELL8 cx Module and Desktop
Computer: 70” W x 30” D x 24” H (180 cm x 75 cm x 60 cm)
Note: Bench space must be capable of supporting 200 pounds (91kg)
Floor space
Waste Container: 8 3/4” W x 14 1/8” H x 6” D (22 cm x 36 cm x 15 cm)
Weight
200 pounds