TA I
NSTRUMENTS
TGA 2050
4–17
Table 4.1
(continued)
Code
Meaning
Initial
The temperature is being
equilibrated to the desired
set point. When the
temperature has reached
equilibrium, the status
will change to “Ready.”
Iso
The thermal program is
holding the current
temperature isothermally.
Jumping
The heater is jumping
ballistically to the set
point temperature.
Load
The TGA is loading a
sample pan onto the
balance.
No Power
No power is being applied
to the heater. Check the
heater fuse.
Opening
The furnace assembly is
opening.
Ready
The system has equili-
brated at the initial
temperature and is
ready to begin the next
segment. Use the Start
function to continue the
method.
(table continued)