TA I
NSTRUMENTS
TGA 2050
3–11
2. Press the
LOAD
key. The TGA will auto-
matically load the sample pan onto the
balance.
3. Position the thermocouple at the edge of the
sample pan, rather than in the middle, for
best results (Figure 3.3).
The position of the thermocouple should be the
same as it was during temperature calibration.
Figure 3.3
Adjusting the
Thermocouple
NOTE:
Sample
Hook
Thermocouple
should be
about 2mm
from the pan
surface.
Sample Pan
Thermocouple
Running an Experiment