maintenance/
troubleshooting
4
2
Synrad
Pulstar p100
operator’s manual
Maintenance
The
Maintenance
section includes subsections:
■
Disabling the p100 laser
■
Daily inspections
■
Storage/shipping
■
Cleaning optical components
■
Output window replacement
Disabling the p100 laser
Before performing any maintenance on your Pulstar p100 laser, be sure to completely disable the laser by
disconnecting the DC power cables from the DC power supply.
Daily inspections
Perform the following steps daily to keep your Pulstar p100 laser in optimum operating condition. Except
for the procedures described below, no other service is required or should be attempted.
Warning
serious
personal
injury
A risk of exposure to toxic elements, like zinc selenide, may result
when certain optical or beam delivery components are damaged. In
the event of damage to laser, marking head, or beam delivery optics,
contact SYNRAD, Inc. or the optics manufacturer for handling
instructions.
Caution
possible
equipment
damage
If you operate your laser or marking head in a dirty or dusty environ-
ment, contact SYNRAD about the risks of doing so and precautions
you can take to increase the longevity of your laser, marking head,
and associated optical components.
1
Inspect all cooling tubing connections for signs of leakage. Check for signs of condensation that may
indicate the cooling water temperature has been set below the dew point temperature. Condensation
will damage electrical and optical components inside the laser. See
Setting coolant temperature
in the
Getting Started chapter for details on preventing condensation.
2
When using compressed air as a purge gas on your p100 laser, empty water traps and oil separators on
each filter and/or dryer between the laser and your compressed air source. Compressed air purity must
meet the purge gas specifications shown in Table 1-3 in the Getting Started chapter.
Summary of Contents for Pulstar P100 Series
Page 2: ......
Page 4: ......
Page 12: ...x Synrad Pulstar p100 operator s manual This page intentionally left blank...
Page 48: ...operation 210 Synrad Pulstar p100 operator s manual This page intentionally left blank...
Page 110: ...index i8 Synrad Pulstar p100 operator s manual This page intentionally left blank...