operator’s manual
Connecting
Cooling tubing connections
To connect cooling tubing to your p100 laser, refer to Figure 1-5 and perform the following steps.
Figure 1-5
Pulstar p100 cooling connections
1
Locate the 1/2-inch O.D. polyethylene cooling tubing in the p100 ship kit.
Note:
If using metric cooling tubing in the application, locate the two 1/2-inch male to 12-mm female
90° adaptors in the cooling kit and install them into the p100’s
Water In
and
Water Out
ports.
2
Cut and connect a length of cooling tubing to fit between the chiller’s Outlet port and the
WATER IN
port on the rear of the p100 laser.
3
Cut and connect a length of cooling tubing to fit between the
WATER OUT
port on the rear of the
laser and the chiller’s Inlet port.
OUTLET
INLET
Chiller
WATER IN
WATER OUT
Caution
possible
equipment
damage
Inlet cooling water temperature must always be maintained above the
dew point to prevent condensation and water damage to your Pulstar
laser.
4
Turn on the chiller and adjust the temperature setpoint to 18 °C to 22 °C. Regulate coolant flow to
5.7–7.6 lpm (1.5–2.0 GPM) at less than 414 kPa (60 PSI) of pressure.
5
Closely examine all cooling connections and verify that there are no leaks.
Summary of Contents for Pulstar P100 Series
Page 2: ......
Page 4: ......
Page 12: ...x Synrad Pulstar p100 operator s manual This page intentionally left blank...
Page 48: ...operation 210 Synrad Pulstar p100 operator s manual This page intentionally left blank...
Page 110: ...index i8 Synrad Pulstar p100 operator s manual This page intentionally left blank...