MAINTENANCE & TROUBLESHOOTING
PAGE
|
95
inputs are recognized. When PWM Command pulses are applied (and are long enough to produce laser
output) the
LASE
LED illuminates red.
On OEM lasers, the
RDY
lamp illuminates on DC power-up after
INT
and
TMP
indicators illuminate green.
The
SHT
LED illuminates blue when a
Shutter Open
request signal is applied. When both
RDY
and
SHT
indicators are lit, the laser is permitted to lase immediately. When PWM Command pulses are applied
(and are long enough to produce laser output) the
LASE
LED illuminates red.
ti Series
RDY
and
SHT
indicators (as well as
Laser Ready
and
Shutter Open
outputs) denote separate
control functions. Although the
RDY
lamp may light while the
SHT
LED is
Off
(Shutter Switch Closed or
Shutter Open Request signal missing), no power is applied to the RF driver until both
RDY
and
SHT
indicators are illuminated.
The following tables show how ti Series LED and output signal status changes as various operating and
fault conditions occur. The
Lase Error Codes
table lists specific laser faults that cause
RDY
and
SH
T
indicators to flash
Laser Fault Indications
ti-Series lasers have the ability to indicate five specific fault conditions. In the event of certain faults, the
RDY
LED will blink an error code, pause four seconds, and then repeat the error code. This sequence
continues until the fault is corrected and the laser is reset by cycling DC power to the laser. If a No-Strike
condition occurs, the
SHT
LED flashes continuously until the gas breaks down into a plasma state.
The table below lists error codes, the corresponding fault condition, and describes possible corrective
actions.
Resetting Faults – Keyswitch Lasers
Remote Interlock Condition - A remote interlock condition occurs when the Remote Interlock input
opens (
INT
LED changes from green to red). To reset a remote interlock fault, re-establish the Remote
Inter-lock signal input (
INT
LED changes from red to green) and cycle the Keyswitch from OFF to ON
(or apply a Remote Reset/Start Request pulse with the Keyswitch set to ON). When the
RDY
LED
illuminates, lasing is enabled after a five-second delay.
Over Temperature Fault - Over temperature faults occur when thermal limits in the laser are exceeded
(
TMP
LED changes from green to red). To reset an over temperature fault, lower coolant temperature
below 30 °C (water-cooled) or 40 °C (air-cooled), cycle DC power to the laser, and then cycle the
Keyswitch from OFF to ON (or apply a Remote Reset/Start Request pulse with the Keyswitch set to
ON). When the
RDY
LED illuminates, lasing is enabled after five-seconds.
Under/Over Voltage Fault - An under-voltage or over-voltage fault occurs when the DC input voltage is
below or above preset limits. This fault is indicated by the
RDY
LED flashing 1 blink (under-voltage) or 2
blinks (over-voltage). To reset an under-voltage or over-voltage fault, ensure that 48 VDC is applied to
the laser under full-load conditions, cycle DC power, and then toggle the Keyswitch from OFF to ON (or
apply a Remote Reset/Start Request pulse with the Keyswitch set to ON). When the
RDY
LED
illuminates, lasing is enabled after a five-second delay.
Summary of Contents for Novanta OEM ti Series
Page 1: ...ENGINEERED BY SYNRAD OEM ti Series User Manual...
Page 26: ...TI60 LABEL LOCATIONS PAGE 26 Ti60 label locations Figure 3 1 Ti60 hazard label locations...
Page 27: ...TI80 LABEL LOCATIONS PAGE 27 Ti80 label locations Figure 3 2 Ti80 hazard label locations...
Page 28: ...TI100 LABEL LOCATIONS PAGE 28 Ti100 label locations Figure 3 3 Ti100 hazard label locations...
Page 32: ...EUROPEAN UNION EU REQUIREMENTS PAGE 32 Table 3 1 Class 4 safety features...
Page 34: ...EUROPEAN UNION EU REQUIREMENTS PAGE 34 Figure 3 4 ti60 Declaration Document...
Page 35: ...EUROPEAN UNION EU REQUIREMENTS PAGE 35 Figure 3 5 ti80 Declaration document...
Page 36: ...EUROPEAN UNION EU REQUIREMENTS PAGE 36 Figure 3 6 ti00 Declaration document...
Page 55: ...INPUT SIGNALS PAGE 55 Table 4 5 Input signal table continued...
Page 65: ...DB 9 PIN DESCRIPTIONS PAGE 65 Table 4 9 Side mounted DB 9 pin descriptions continued...
Page 73: ...REMOTE INTERLOCK FUNCTIONS PAGE 73 Table 4 10 Ti60 general specifications continued...
Page 74: ...REMOTE INTERLOCK FUNCTIONS PAGE 74 Table 4 11 Ti80 general specifications...
Page 75: ...REMOTE INTERLOCK FUNCTIONS PAGE 75 Table 4 11 Ti80 general specifications continued...
Page 76: ...REMOTE INTERLOCK FUNCTIONS PAGE 76 Table 4 12 Ti100 general specifications...
Page 77: ...REMOTE INTERLOCK FUNCTIONS PAGE 77 Table 4 12 Ti100 general specifications continued...
Page 82: ...TECHNICAL DRAWINGS PAGE 82 Figure 4 29 Ti Series HS Outline and Mounting...
Page 83: ...TECHNICAL DRAWINGS PAGE 83 Figure 4 28 Ti 100p Outline and Mounting...
Page 84: ...TECHNICAL DRAWINGS PAGE 84 Figure 4 29 Ti 100 Fan Outline and Mounting...
Page 85: ...TECHNICAL DRAWINGS PAGE 85 Figure 4 30 ti Series packaging instructions...
Page 93: ...MAINTENANCE TROUBLESHOOTING PAGE 93 Table 4 14 Status Signals...
Page 99: ...MAINTENANCE TROUBLESHOOTING PAGE 99...
Page 105: ...APPENDIX PAGE 3 This page is intentionally left blank...