Press calibrate to validate the entry.
-
In case of correct calibration "Calibration OK" is displayed in green
-
In case of calibration error "Error Calibration" is displayed in red
➢
Reset to factory calibration
Select the Erase tab.
Press Clear.
-
"Calibration OK" information is displayed in green to confirm the
deletion.
Warning: this operation is irreversible. Once confirmed by the Clear key,
your calibration parameters for this parameter will be lost.
b)
Chlorine/Bromine
Pressing the « Calibration » button opens the programming window.
Offset
Gain
Erasing
Before calibrating the chlorine measurement of the CU / PT probe, make sure that :
-
The pH is stabilized at its nominal operating value of your installation
-
The Chlorine or Bromine is also stabilized at the nominal operating value of your installation.
➢
Zero calibration of the probe
-
Close the water circulation in the CU/PT chamber
-
Wait for the measure stabilization, minimum 5min
-
Enter the calibration menu:
o
Offset part
o
Enter the 0ppm value as the standard value
o
Press Calibrate
Calibrate
Clear
Clear
Settings
Chlorine channel
Calibration
Calibration type selection
Measured value
0.04
Enter the standard value
0.04
Calibrate
Offset Gain Efface
Settings
Chlorine channel
Calibration
Calibration type selection
Measured value
1.50
Enter the standard value
1.50
Calibrate
Offset Gain Efface
Settings
Chlorine channel
Calibration
Calibration type selection
Factory calibration
Will be restored
Clear
Offset Gain Efface