General use
Page 39/52
Installation, commissioning and operating instructions
SYCLOPE
TERE’O Touch
®
CU/PT
f)
Maintenance
Pressing the « Maintenance » button opens the programming window.
➢
POT measurement input:
This input corresponds to the pH measurement input. The raw measurement value
(in mV) can be read, the pH value corresponding to the adjusted measurement of the
calibration. The Offset and Gain information correspond to the calibration of the sensor (0mV &
100% correspond to an uncalibrated sensor).
➢
IN1 measurement input:
This input corresponds to the measurement input of the temperature. You can read
the raw measurement value (in mA), the value of the temperature corresponding to the
adjusted calibration measurement. The Offset and Gain information correspond to the
calibration of the sensor (0mA & 100% correspond to an uncalibrated sensor).
➢
AUX measurement input:
This input corresponds to the measurement input of Chlorine or Bromine. The gross
measurement value (in µA), the value of Chlorine or Bromine corresponding to the adjusted
measurement of the calibration can be read. The Offset and Gain information correspond to the
calibration of the sensor (0%A & 100% correspond to an uncalibrated sensor).
➢
K1 Dry contact input:
This entry corresponds to the entry dedicated to the tank level pH channel.
Depending on the configuration of the input you can see its status and type of configuration.
➢
K2 Dry contact input:
This entry corresponds to the entry dedicated to the tank level chlorine/Bromine
channel. Depending on the configuration of the input you can see its status and type of
configuration.
➢
FLW Dry contact input :
This input corresponds to the entrance dedicated to the flow of water in the
measuring chamber. Depending on the configuration of the input you can see its status and
type of configuration.
This screen is constantly updated.
Maintenance
- pH
- Temperature
- Chlorine
- Flow-Switch
- Offset : 0mV
–
Gain : 100%
- Offset : 0.0mA
–
Gain : 100%
- Offset : -9.0µA
–
Gain : 100%
Inactive Active - NC
Not used
Active Active - NC
-10
mV
7.18
pH
13.0
mA
23.26
°C
10.5
µA
0.04
ppm
- pH tank level
- Chlorine tank
level