The detector should be kept under vacuum all the time.
The best long term storage environment is an oil free vacuum in the 1E-2mbar range or
better. The transport container is not designed for a long term storage of the detector. A
storage time of one month should not be exceeded.
The transport housing should be evacuated regularly with a period of at least one month,
in case that the transport housing must be used for a longer storage time.
Note
3.2.2 Detector Orientation
The red dots in
and
mark the 0/0 position of the DLD image which corresponds to the upper
left corner of the image in the software.
Figure 1a: 0/0 position of the DLD R2.55 image (red dot).
Image 0/0 position
Figure 1b: 0/0 position of the DLD R3.54 image (red dot).
Image 0/0 position