Delayline Detector DLD4040 Manual
DLD4040 R2.55 & R3.54 Manual | Surface Concept GmbH
3 Installation
3.1 Initial Inspection
Visual inspection of the system is required to ensure that no damage has occurred during shipping. Should
there be any signs of damage, please contact SURFACE CONCEPT immediately. Please check the delivery
according to the packing list (see
) for completeness.
• Delayline Detector Unit (delivered in a vacuum container)
• 1x Pulse Processing Unit ACU
• 1x SHV Termination Plug
• 1x DLD Readout Cable (HDMI) with Strain Relief
• Storage Medium with Documentation and Software
Table 1: Packing list for the delayline detector.
3.2 Installation
3.2.1 Mounting the Delayline Detector
The detector is transported under vacuum. Proceed as follows, to install it into your vacuum chamber:
• Vent the transport container
carefully and slowly (use N2 for venting if possible)
.
• Remove the transport container
carefully
.
• Before installing the detector to your chamber, check the front side of the MCP stack for particles.
The microchannel plates in front of the detector should be protected from exposure to
particle contamination. Particles that stick to the plate can be removed by using a single-
hair brush carefully and/or with dry nitrogen. Reading the instructions “microchannel
plates” in
is strongly recommended.
• Install the detector into your vacuum chamber.
• Keep the vacuum container in case that the detector must be sent back for repair. It can also be used
to store the detector when not installed in a vacuum chamber.