27
Delayline Detector DLD4040 Manual
DLD4040 R2.55 & R3.54 Manual | Surface Concept GmbH
Figure 11: Internal connection of high voltage potentials (schematic).
The resistance between “
MCP-B
” and “
MCP-F
” (resistance of MCP stack) should be
in the range of 8 – 30 MΩ (the exact value is given in the specification sheet of your
detector), while the resistance between “
MCP-B
” and “
U_DLD
” and “
MCP-F
” and “
U_DLD
”
respectively should be unlimited. This can be used to verify the position of the “
U_DLD
”
voltage connector.
Note