39
Delayline Detector DLD4040 Manual
10 Technical Data
Base Flange R2.55/ R3.54:
Active area of the DLD 4040:
Operation voltage at detector:
Voltage at MCP front:
Voltage difference between MCP back and anode:
Potential of Detector Cover:
Grid voltage:
Signal Feedthrough:
HV Feedthrough R2.55/ R3.54:
Bake-out temperature (max.):
Vacuum pressure range for operation:
Delayline Detector General
CF160/ CF100
Ø 40mm round
see specification sheet
see specification sheet
typ. 400 V
connected to MCP-F potential
not available
1x 4-fold SMB Feedthrough on CF40 Flange
3x SHV5 Feedthrough on CF40 Flange/
3x SHV5 Feedthrough on CF16 Flanges
150°C
< 1E-6 mbar
No. of Amplifier-CFD Channels:
Bandwidth of DLD amplifiers:
CFD working frequency:
CFD jitter (max.):
CFD walk (typ.):
Amplifier CFD - Unit ACU 3.4.2
4
1.6 GHz
200 MHz
20ps
< 50ps (while ambient temperature varies less then 5 K)