![SUPMEA SUP-1158-J User Manual Download Page 55](http://html1.mh-extra.com/html/supmea/sup-1158-j/sup-1158-j_user-manual_1406088055.webp)
Chapter 6 Manual Zero Point Windows display codes
www.supmea.com
- 46 -
Temperature sensitivity and user SHC
When the delta temperature is less than
the sensitivity set,, energy will not be
accumulated. Set the adjustable
temperature range of 0
℃
to 10
℃
.
The factory default setting is 0.2
℃
When the user specific heat is setting,
energy is accumulated according to the
user specific heat value. The setting
range is from 1 to 99999 KJ/m3 C.The
factory default is 4186.8KJ/m3C.
Energy Totalizer Switch
Select "ON" represent to open Energy
Totalizer; Select "OFF" represent to
close Energy Totalizer.
Energy Multiplier
Select Energy Multiplier range:
10-3 ~ 104 (E-3 ~ E4)
Reset Energy Totalizer
Select "YES" to reset Energy
Totalizer value.
Reset Ener gy Total
N
O
Ener gy Totali zer
[87 ON
Temperatur e Delicacy
0.20 C
Energy Multiplier [88
3.
x1
=(EO)