1. Specifications
Model
TC-7310-RS-M
Measuring Mode
MLSS
Measuring Unit
mg/l, ppm
Measuring Range
0~50,000 mg/l (ppm)
(TS-MxS-A sensor)
Resolution 1
mg/l
(ppm)
Calibration Matrix Type
Selection by MLSS, Defined, Digested Sludge and Activated Sludge.
Except MLSS, other 3 matrix types provide display value alarm
function when the measuring value exceeds the defined boundary value
Calibration Mode
MLSS solution up to 5 points calibration
and calibration by factor correction
Electrode Zero
Sensor internal zero point correction activation function
Ambient Temp.
0~50
C
(TS-MxS-A Sensor working temp.: 0~40
C)
Storage Temp.
-20~70
C
Display
Large LCM with sensor for backlight and contrast
Text mode: Numerical display
Chart mode: 3 min real-time dynamic graph
Trace mode: Programmable 3 min to 4 week trend graph
Analog Output
Isolated DC 0/4~20 mA corresponding to main measurement,
Max. load 500
Serial Interface
Isolated RS-485 (MODBUS RTU or ASCII)
Logbook 50
events
Settings
Contact
RELAY contact, 240 VAC 0.5 A Max. (recommended)
Activate
Hi/Lo, Hi/Hi, Lo/Lo selectable, two limited programmable,
ON/OFF
Wash
RELAY contact: ON 0~99 min 59 sec / OFF 0~999 hr 59 min
Sensor Wiper
Auto, sensor internal function for fixed 10 min/per wipe
Manual, setting from 2~9 min/per wipe
Voltage Output
DC
±
24V
Power Supply
100 V~240 VAC ±10 %, 8W max., 50/60 Hz
Installation
Wall or Pipe or Panel Mounting
Dimensions
144 mm x 144 mm x 115 mm (H
W
D), 1/2 DIN
Cut-off Dimensions
138 mm
138 mm (H
W)
Weight 0.8
kg
Protection
IP65 (NEMA 4X)
Note: The specifications are subject to change without notice.
4