5. Operation
5.1 Measurement Mode:
After all electrical connections are secured and tested, connect the instrument to the power
supply and turn it on. The transmitter will automatically enter measurement mode with the
factory default settings or the previous user settings.
5.2 Setup Menu:
Please refer to the set-up instructions in Chapter 7. Press and simultaneously to enter
set-up menu, and press to
return to measurement
mode.
5.3 Calibration Menu:
Please refer to the calibration instructions in Chapter 8. Press and simultaneously to
enter calibration menu, and press to return to measurement
mode.
5.4 Shortcuts:
1. Under measurement mode, press and hold for 2 seconds to enter Logbook function.
Press to return to measurement mode.
2. Under measurement mode, press and hold for 2 seconds to switch from display mode to
text mode, trace mode, or real-time chart display mode.
5.5 Default Values:
5.5.1 Settings Default Values:
Sensor model: TS-MxS-A
Matrix table: MLSS
Measuring unit: mg/l
Product adjustment: 0 mg/l
Relay 1: High point alarm: AUTO, SP1=40,000 mg/l, Hys = 10 mg/l
Relay 2: Low point alarm: AUTO, SP2= 10,000 mg/l, Hys = 10 mg/l
Wash time: OFF
Wiper: Manual, 9 minute/per wipe
Analog current output: 4~20 mA, 0~50,000 mg/l
Date & Time: 2016/1/1 00:00:00
RS-485: RTU, 19200, Even, 1, ID: 001
Digital filter: 30
Backlight settings: OFF
Contrast: 0
Logbook: None
Auto return: Auto, 3 minutes
Code settings: OFF
11