![SunTech 99-0133-00 Manual Download Page 26](http://html1.mh-extra.com/html/suntech/99-0133-00/99-0133-00_manual_1400832026.webp)
UI Item
Description
Minimum Reading
Requirements:
Enter the requirements for a minimum percentage of scheduled readings captured
during the defined QC period.
The minimum number of readings captured per hour during the defined QC period
(between 1 and 12).
The minimum number of hours to contain readings for the entire study.
NOTE: To access the Quality Control tab you must first turn on the “Quality Control”
dial at the bottom of the Program Monitor window.
Fitting a Patient with the Oscar 2 and ABPM Cuff
After you have successfully programmed the Oscar 2 using AccuWin Pro 4, you may begin fitting the patient
with the monitor and a blood pressure cuff. Cuffs may be used on either arm.
1.
Choose the proper cuff size
To determine the correct cuff size for your patient, wrap the cuff around the patient’s upper arm without
sliding the arm through the sleeve. Use the color- coded
RANGE
indicator on the inside of the cuff and the
bold
INDEX
marker to check that the arm circumference falls within the cuff range. If the arm is within range,
this cuff size is correct for your patient. If the marker is outside the
RANGE
indicator, select a new cuff size as
indicated by color.
CAUTION:
Using an incorrect cuff size could result in erroneous and misleading blood pressure measure
-
ments.
2.
Apply the ABPM cuff
To apply the ABPM cuff, simply slide the sleeve up the patient’s arm, ensuring the color size indicator is at the
top of the cuff. The cuff should be midway between the elbow and shoulder. Be sure the ARTERY indicator
is over the patient’s brachial artery, between the bicep and tricep muscles. Wrap the cuff snugly around the
patient’s upper arm. There should be space for approximately 1 finger underneath the bottom of the cuff.
Figure 6: Instructions for Applying the ABPM Cuff
3.
Connect the BP hose to the cuff
Connect the BP hose to the cuff by pushing the fittings together and twisting until it is locked into place. The
hose can be released by twisting the connectors and pulling the female fitting away from the male connec
-
tion. Drape the hose over the patient’s shoulder, behind the neck and across to the opposite side of the body.
4.
Attach to patient
Page 26 of 79
80-0062-00-RevL 2020-9-14