12. Overview of AccuWin Pro 4 Report Symbols
The following symbols are used in the Reviewed and Omitted BP data report pages.
Symbol
Explanation
+
Signifies the reading was initiated manually by the patient pressing the Start/Stop key
on the monitor.
-
Indicates that the pressure to which the cuff inflated in dynamic inflation mode may
have been too low to obtain an accurate systolic reading. Readings marked with “-” are
automatically omitted from the study. Compare the tagged reading with a “-” to the BP
readings above and below it. If the omitted reading seems to be acceptable, un-tag it
from the ABP Data tab; if it does not seem acceptable, leave it “tagged”.
r
Indicates that this reading was a retry reading. Retries are readings that are
automatically obtained 4 minutes after a measurement attempt fails or a questionable
reading is obtained by the monitor.
!
The exclamation point is a permanent tag and is used when an accurate reading
cannot be determined by the Oscar 2.
>
This symbol will appear to the left of the measurement if the reading results were
above the set threshold.
r01 to
r30
These numbered tags identify a series of patient-marked events which occurred during
the study.