PULSE 6 SERVICE MANUAL
PAGE 7.7
2009
SUNRISE MEDICAL
Section 7
Shock Removal
The shocks used in this power base are pre-
adjusted at the factory. Contact Sunrise Medical
Technical Service if adjustment is required.
1.
Remove the batteries. Refer to the Battery
Removal section of this manual.
2.
Remove the motor/gearbox assembly from
the side on which you are removing the shock.
Refer to Motor/Gearbox Removal section of this
manual.
3.
Use a 17 mm wrench and remove the top
bolt securing the shock to the base (
fi
g. 7.20).
4.
Use two 19 mm wrenches and remove
the bottom nut and bolt securing the shock to the
base (
fi
g. 7.21 & 7.22).
5.
Lift the shock from the base.
6.
Reverse the previous procedures to install
the shock.
Removal Procedures (cont.)
fig. 7.20
fig. 7.21
fig. 7.22