PULSE 6 SERVICE MANUAL
PAGE 4.2
2009
SUNRISE MEDICAL
Section 4
5.
Disconnect the power connector from the
control module, and use a Multimeter to check
for battery voltage at the connector (fig4.3). If
voltage is present, replace the control module.
Retest as necessary.
Note: The power connector is the larger 2-pin
connector between the left and right motor
cable.
6.
Manually tilt the seat back and remove the
shroud from the base. Refer to Battery Removal
section of this manual.
7.
Verify that the circuit breaker located at
the front of the power chair is not tripped (fig 4.4).
Reset if it is tripped and retest.
8.
Disconnect the batteries (fig 4.5).
Rnet Troubleshooting Procedures
fig. 4.3
fig. 4.4
fig. 4.5