PULSE 6 SERVICE MANUAL
PAGE 5.8
2009
SUNRISE MEDICAL
Warning: Use caution in the next step. Do not short the leads of the Multimeter together
or damage may occur to the control module.
6.
At the 6-pin connector on the cable leading
to the control module, verify that battery voltage
is present between the indicated pins (
fi
g. 5.12).
Note: The polarity of the voltage is not important
since it is reverses when the opposite direction is
selected.
7.
If voltage is not present, replace the cable
leading to the control module. If this does not
solve the problem, replace the control module.
Refer to Control Module Removal section of this
manual. Retest as necessary.
Voltage to
the actuator
8.
At the 6 pin connector (
fi
g. 5.13) on the rear
of the tilt, check continuity between the indicated
pins (
fi
g. 5.14).
R-net Fault Codes cont.
fig. 5.12
fig. 5.13
fig. 5.14
Section 5