PULSE 6 SERVICE MANUAL
PAGE 5.5
2009
SUNRISE MEDICAL
3.
Locate the creep micro-switch mounted at
the rear of the actuator (
fi
g. 5.3).
Note: The creep micro-switch may be
seen from the top of the seating system on the
right rear of the actuator. It may be necessary to
remove the seat pan to access the switch.
4.
Manually operate the creep micro-switch
while observing the hand control or Omni for the
turtle symbol to go on and go off with the micro-
switch operation. If the symbol changes, the
problem is in the programming. Contact Sunrise
Medical Technical Service for assistance. If the
symbol does not change, proceed to step 5.
5.
Lower the tilt system.
6.
Locate the 6-pin connector at the rear of
the tilt (
fi
g. 5.4).
7.
Disconnect the connector leading to the
control module.
8.
Test the micro-switch wiring by performing
the following test.
a.
Ensure that the seat is in the full down
position.
b.
Use Multimeter and measure continuity
between the indicated pins below on the 6-pin
connector mounted on the tilt. When down the
indication should be open (
fi
g. 5.5).
c.
If it is closed, replace the micro-switch
assembly. Refer to Tilt Actuator and Micro-Switch
Removal section of this manual.
d.
Retest as necessary.
9.
If the above test passes, replace in this
order:
a.
Cable leading from tilt to control module.
b. Control
module
10.
Retest as necessary
Open
R-net Fault Codes cont.
Section 5
fig. 5.5
fig. 5.4
fig. 5.3