PULSE 6 SERVICE MANUAL
PAGE 4.1
2009
SUNRISE MEDICAL
Section 4
R-net Troubleshooting Procedures
Chair Will Not Power Up
2
1
3
1.
Check for battery voltage at the hand
control using a Multimeter. Connect the Multimeter
between the two outside pins (fig 4.1), pin 1 is
positive (+), pin 2 is negative (-). Note: Positive
is on the right.
2.
If battery voltage is present, replace the
following components in this order:
a. Cables
b. Hand
control
c.
Control module. Refer to Control Module
Replacement section of this manual.
d.
Retest as necessary.
3.
Verify that the buss cables are correctly
mated between the hand control and the control
module (fig 4.2).
Note: This connector is mated incorrectly.
The connectors are designed to visually
indicate when they are not mated correctly.
If yellow is showing between the halves (A),
push them closer together until only black is
visible. Retest as necessary.
A
fig. 4.1
fig. 4.2