![Sumitomo TYPE-71c Manual To Operation Download Page 44](http://html1.mh-extra.com/html/sumitomo/type-71c/type-71c_manual-to-operation_1390693044.webp)
2-19
MENU
SM G652 Std.
60mm 0.25
X/Y
Tool
DATA
DISP
Reset
GO
Estimation Loss 0.01dB
Tap Go icon
MENU
SM G652 Std.
60mm 0.25
X/Y
Tool
DATA
DISP
Estimation Loss 0.01dB
Reset
GO
Tap Go icon
MENU
SM G652 Std.
60mm 0.25
X/Y
Tool
DATA
Estimation Loss 0.01dB
Tap Go icon
Reset
GO
DISP
Evaluating splice quality
If the spliced fiber is as follows or estimated losses are high, re-do the entire splice.
Bubble
Black line
(Re-do splicing.)
(Re-do splicing.)
Bulging
Neck-down
(Re-do splicing.)
(Re-do splicing)
Kink or offset
White line
(Re-do splicing.)
(Re-arc
→
NG
→
Re-do splicing)
After MMF or dissimilar fiber splicing, a line sometimes appears at the
splice point. But it does not affect splice quality, such as splice loss,
tensile strength, and etc.