XT
-
VISION
F
LOW RATE ACTUATORS
B3.3
N
AME
:
T
RAINING DEPARTMENT
D
ATE
:
04/2018
P
AGE
:
52/109
REFERENCE:
TECHNICIAN’S
OPERATING
MANUAL
POLYVRAC
XT
VISION
I
ND
02
At the end of the cyvle, read the
V SHUT
and
V OPEN
values on
Channel 1
and
Channel 2
For
V SHUT
, only keep the lowest value of
Channel 1
and
Channel 2
and deduct
0.03 from it
Example: Channel 1-V SHUT = 6.01 and Channel 2-V SHUT = 5 .96
then: 5.96 – 0.03 = 5.93
Enter the new
V SHUT
value on
Channel 1
and
Channel 2
For
V OPEN
, only keep the highest value of
Channel 1
and
Channel 2
and add
0.03 to it
Example: Channel 1-V OPEN = 0.02 and Channel 2-V OPEN = 0.03
then: 0.06 + 0.03 = 0.06
Enter the new
V OPEN
value on
Channel 1
and
Channel 2
Press
ESC
to exit
Diagnostic on the console:
Access the
SETUP
page with the
key
Press
for
"Diagnostics"
Select
In this page, you can control the electric flow rate actuators manually and
check their position.