Chapter 4
The
GO
BED Maintenance Manual
42
Head Section Replacement
Bed without Optional CPR
1. Raise the bed to high position, lock casters and bring mattress support to horizontal
position.
2. Remove the head board.
3. Raise Fowler to high position.
4. Unplug the power cord from the wall receptacle.
5. Remove both head safety side-rails. Refer to the "Head Safety Side-Rail Replacement"
procedure, section 4.1, page 24)
.
6. Remove the two mattress side retainers (
A
) and save them for the replacement head
section.
7. Plug the bed power cord and bring Fowler back to horizontal position.
8. Remove the Rue ring/washer/nylon washers(2)/clevis pin (
G
, fig. 4.14, page 44) hooking up
the head actuator tube to the head section lever arms.
N
N
O
O
T
T
E
E
Apply grease on the clevis pin and the nylon washers before hooking up the actuator tube to the
head section.
9. Manually lift the head section completely and attach it to the foot section using a ratchet tie-
down strap to stabilize it.
Figure 4.13D
TÊTE
A
CENTRE SECTION
HEAD
END
HEAD SECTION
D
D
C
B