Chapter 3
The
GO
BED Maintenance Manual
22
PROBLEM
POSSIBLE CAUSES
RECOMMENDED ACTION
The micro switch cable
improperly connected to the
PC Board.
Check cable connection
(see section 4.8). Refer to
Connection Diagram in
Appendix A for proper
connecting position.
The micro switch cable running
from the micro switch to the
PC Board connector defective.
Replace micro switch
cable.
Head or Knee Gatch actuator
defective.
Replace head or Gatch
actuator (see section 4.14).
PC Board defective.
Replace PC Board
(see section 4.8).
Fowler and Knee Gatch are
out of synchronization when
optional Contour Positioning
is activated.
Micro switches detecting the
relative positions of the
mattress support sections not
functioning properly.
Check that micro switches
and micro switch activator
(located on the left side of
the fixed frame) are tightly
fastened.
Check micro switches and
wires connected to them for
proper connection. Replace
component as needed (see
section 4.18).
Micro switch cable improperly
connected to the PC Board.
Check cable connection
(see section 4.8). Refer to
Connection Diagram in
Appendix A for proper
connecting position.
Micro switch cable running
from the micro switch to the
PC Board defective
Replace micro switch
cable.
Head or Gatch actuator
defective.
Replace head or Gatch
actuator (see section 4.14)
PC Board defective.
Replace PC Board
(see section 4.8)
Nurse call command does
not function.
Nurse call wall connection
cable plugged into a defective
wall receptacle.
Check wall receptacle.
Nurse call wall connection
cable disconnected or
improperly connected to the
multiple connector.
Check cable wire
connection to the multiple
connector (see section 4.5).
Nurse call wall connection
cable severed or defective.
Replace nurse call wall
connection cable (see
section 4.5)