Maintenance Procedures
Chapter 4
39
10. Remove the screw (
C
) holding the duplex receptacle (
D
) to the housing.
11.
Loosen the screws holding the wires to the duplex receptacle. Note their connecting
positions.
12. Test the duplex receptacle.
- if defective, proceed with step 13 through step 15 and end procedure.
- if functioning properly, go to step 16.
13. Replace defective duplex receptacle by one of the same type (Leviton 8200-W).
14. Reverse above steps to reassemble auxiliary outlet with replacement duplex receptacle.
15. Test auxiliary outlet for proper operation before returning bed to service.
End of procedure.
16. Disconnect auxiliary outlet cable (
E
) from PC Board.
17. Remove locknut (
F
) holding cable ground wire (
G
) to the PC Board housing and remove
outlet cable. Note the position of the ground wire among the other.
N
N
O
O
T
T
E
E
Replace the auxiliary outlet cable ground wire exactly at the same position it was before being
removed.
18. Test the auxiliary outlet cable:
- if defective, proceed with step 19 through 21.
- if functioning properly, suspect a problem with the PC Board (refer to section 4.8 for the PC
Board Replacement procedure).
19. Replace defective auxiliary outlet cable.
20. Reverse above steps to reassemble auxiliary outlet with replacement cable.
21. Test auxiliary outlet for proper operation before returning bed to service.
4.13 MATTRESS SUPPORT SECTION REPLACEMENT
Foot Section Replacement
1. Raise the bed to high position and lock casters.
2. Put the head side-rails in the down position and raise the foot side-rails.
3. Raise the Knee Gatch to maximum height.
Figure 4.13A
TÊTE
KNEE GATCH SECTION
D
D
A
B
C
A
FOOT SECTION
HEAD
END
Summary of Contents for The Go BED FL17E