Chapter 2
The
GO
BED Maintenance Manual
14
2.2 LUBRICATION
Listed below are the lubrication points and their recommended time interval check. When
needed, lubricate these points with OG2 grease (Stryker Bertec part number M0027).
The use of types of grease other than the one recommended (OG2 grease) could lead to
deterioration of critical parts and to mechanism failure, resulting in injury to the patient or user
and damage to the bed.
The
GO
BED uses oil-impregnated shoulder spacers at hinge points.
Do not
lubricate these
shoulder spacers. When shoulder spacers are found worn, replace them.
Annual Checklist:
q
The Hi-lo lever nylon sliders and their shafts (see fig. 2.2A, page 15). Check that grease is
still present along the course of the sliders (inner surfaces (bottom and side) of the rail).
q
Lower mounting points (see fig. 2.2A, page 15) of the lifting system stabilizer.
q
Side-rail plungers and plunger springs (see fig. 2.2B, page 15).
q
The side-rail arm glide rods (see fig. 2.2B, page 15).
Two Years Interval Checklist:
q
The two Hi-lo actuator tubes (see fig. 2.2C, page 15) to facilitate their sliding into the
support tubes. Bring the bed to its highest position before greasing the tubes.
q
Head, foot and Hi-lo actuator screws (see fig. 2.2C, page 15).
q
Head actuator screw guide (see fig. 2.2C, page 15).
q
Clevis pins and nylon washers linking the actuator tubes to the head and Knee Gatch
section lever arms and the two Hi-lo lever arms (see fig. 2.2C, page 15).
q
Actuator bolts that hold actuators to their brackets as well as the inner sides of the brackets,
including the pivot pin (see fig. 2.2C, page 15).
q
Micro-switch activator (see fig. 2.2E, page 15) of the optional Auto Contour mechanism.
Five Years Interval Checklist:
q
The side-rail shafts and transfer plate handles (see fig. 2.2B, page 15).
q
Inside of the Hi-lo lever moulded bearings (see fig. 2.2A, page 15).
q
The damper bearings, the pivot shaft of the CPR support, and the activation rod pivot sleeve
(see fig. 2.2D, page 15) all from the instant CPR release mechanism.
Serial Number:_______________
_______________
_______________
_______________
_______________
_______________
_______________
_______________
_______________
_______________
_______________
_______________
Completed by:______________________________ Date:_______________
C
C
A
A
U
U
T
T
I
I
O
O
N
N
W
W
A
A
R
R
N
N
I
I
N
N
G
G
Summary of Contents for The Go BED FL17E
Page 2: ......